measuring tools for radiationX

Ecotest TERRA Dosimeter МКS-05

Ecotest TERRA Dosimeter МКS-05

Ingediend door toon op vr, 07/14/2023 - 10:05

Features

  • Big display with luminescent backlight

  • Simultaneous indication of units of measurement, measurement errors, threshold level and real time on the display

  • Analog ten-segment indicator of registered radiation intensity

  • Ability to perform measurements with a preset error

  • Four-level indication of battery discharge

  • Built-in memory which allows storing up to 1 200 measurements

  • Mode of PC connection via Bluetooth

  • Five independent measuring channels with alternate indication of data on the single liquid crystal display

  • Built-in gamma-beta-sensitive Geiger-Muller counter

  • Prompt evaluation of gamma background within 10 seconds

  • Automatic subtraction of gamma background at measurement of beta contamination

  • Measurement results averaging, manually and automatically interrupted

  • Automatic setting of measurement intervals and ranges

  • Audio, vibration and vibration-audio alarm of each detected gamma-quantum or beta-particle that can be switched off

  • Two-tone audio, vibration and vibration-audio alarm of exceeded programmed threshold levels

  • Clock and alarm clock

Specifications

Measurement range of photon-ionizing radiation DER 0.1 … 9 999 µSv/h
Main relative permissible error limit of photon-ionizing radiation DER measurement with confidence probability of 0.95 (15+2/H*(10)) %,where H*(10) is a numeric value of the measured DER in μSv/h
Measurement range of photon-ionizing radiation DE 0.001 … 9 999 mSv 
Main relative permissible error limit of photon-ionizing radiation DE measurement with confidence probability of 0.95 ± 15 %
Energy range of registered photon-ionizing radiation 0.05 … 3 MeV
Energy dependence when measuring gamma radiation DER and DE relative to 0.662 MeV energy (137Cs) in the energy range from 0.05 to 1.25 MeV, not more  ± 25 %
Measurement range of surface beta-particles flux density 10 … 100 000 part./(cm²·min) 
Main relative permissible error limit of beta-particles flux density measurement with confidence probability of 0.95 (20+200/φβ) %, where φβ is a numeric value of the measured surface flux density in part./(cm²·min)
Energy range of registered beta-particles 0.5 … 3 MeV
Measurement range of DE accumulation time 9 999 h
Resolution of DE accumulation time display in the range- of 0 h to 100 h

 

– of 100 h to 9 999 h

1 min

 

1 h

Absolute permissible error limit at DE accumulation time measurement during 24 h ± 1 min
Time of the dosimeter operating mode setting, not more 1 min
Battery life (AAAx2 of 1 280 mAh capacity) under + 20 °С temperature, natural background radiation and switched off display backlight, not more 2 000 h
Operating supply voltage 3 V
Operating temperature range – 20 … + 50 °С
Dimensions, not more 55 × 26 × 120 mm
Weight, not more 0.2 kg