PM1610 series of electronic personal dosimeters (EPDs) are intended for measurement of the personal dose equivalent Hp(10) and personal dose equivalent rate Ḣp(10). The dosimeters are suitable for multiple applications providing the measurement of X-ray (continuous and pulsed) and gamma radiation in the wide energy range.
PM1610 dosimeters have unique features for operation in workplaces requiring the use of personal protective equipment or a harsh environment, including a shockproof rubberized case, a high contrast display with a fluorescent backlight, and two big buttons for easy use even while wearing protective gloves.
Detector |
Geiger-Mueller tube |
Dose indication range |
0.001 µSv to 12 Sv |
Dose measurement range |
0.05 µSv to 10 Sv (continuous photon radiation) 10 µSv to 10 Sv (pulsed photon radiation) |
Dose measurement accuracy |
± 20 % |
Dose rate indication range |
0.01 µSv/h to 12 Sv/h |
Dose rate measurement accuracy |
±(15+0.0015/Ḣ) % (where Ḣ is the measured dose rate in mSv/h) |
Minimum pulse duration of pulsed X-ray radiation |
1 ms |
Energy range |
20 keV to 10 MeV |
Energy response relative to 0.662 MeV (137Cs) |
−60 % (20 keV to 33 keV) −40 % (33 keV to 48 keV) ±30 % (48 keV to 3 MeV) ±50 % (3 MeV to 10 MeV) |
Alarms |
visual, audible, vibration |
Thresholds |
2 independent thresholds for both dose and dose rate |
Memory |
7500 events |
PC communication |
USB |
Power supply |
built-in rechargeable battery |
Battery lifetime |
650 hours* *when the average dose rate is not more than 0.3 µSv/h; backlight and alarms are activated for not more than 20 s per 24 hours |
Ingress protection |
IP65 |
Drop test |
1.5 m onto hard wooden surface |
Dimensions |
58 × 59 × 20 mm |
Weight |
70 g |
Operating temperature |
–20 °С to 50 °С |
Atmospheric pressure |
84 to 106.7 kPa |
Humidity |
98 % at 35 °С |
Mean recovery time |
60 min |
Mean service lifetime |
10 years |
READ BROCHURE